MODULO: TECNICHE DI ANALISI Spettroscopie Vibrazionali Infra-rossa e Raman C. Daniel Sommario • Spettroscopia IR – – Principio della tecnica Aspetti strumentali • • • – Metodi di campionamento • • • Funzionamento spettrometro Componenti Parametri di misura Gas, liquido, solido Trasmissione, Riflettanza Spettroscopia Raman – – Principio della tecnica Alcuni aspetti strumentali Why IR spectroscopy is so widely used? 1. 2. 3. 4. 5. 6. 7. 8. VERY POWERFUL TECHNIQUE SPECTRA ARE (RELATIVELY) EASY TO INTERPRET OPERATION OF THE EQUIPMENT IS EASY ALL TYPES OF SAMPLES CAN BE ANALYZED (GAS, LIQUID, SOLID) FAST SPECTRUM ACQUISITION NO CALIBRATION ALMOST NO MAINTENANCE LOW COST OF BASIC SPECTROMETERS (15000 euros) IR spectroscopy Measurement of the amount of radiation absorbed by a material within IR region IR spectrum wavelength 0.8-1000 µm Frequency: 12800 - 10 cm-1 Divisions arise because of different optical materials and instrumentation Energia delle molecole Come nel caso degli atomi, anche le molecole hanno una energia interna dovuta ai diversi contributi: Emolecola= Eelettronica + Erotazionale+ Evibrazionale + Etraslazionale + Enucleare Eelettronica è associata agli elettroni Enucleare è associata al nucleo Etraslazionale è associata al movimento traslazionale della molecola (3/2 kT) Erotazionale è associata alla rotazione della molecola su se stessa (intorno a uno o più assi), Evibrazionale è associata al movimento reciproco degli atomi nella molecola Etotale = Eelettronica +Erotazionale+Evibrazionale L’energia delle molecole (come quella degli atomi) è quantizzata, per cui sono “permessi” solo determinati livelli di energia Stati elettronici Stati vibrazionali Stati rotazionali Rappresentazione schematica dei livelli energetici molecolari. A ogni livello elettronico sono associati diversi livelli vibrazionali e a ciascuno di questi, a sua volta, diversi livelli rotazionali Frequenza di assorbimento Si descrive una molecola biatomica come un sistema di due masse collegate da una molla Ep = ½ k · x 2 = 2π2µνoscix2 Ep= Energia potenziale k= Costante di forza x = Allungamento µ= Massa ridotta = m1. m2 / (m1 + m2) νosci= Frequenza dell’oscillatore ν osci 1 = 2π k µ νosc direttamente proporzionale alla costante di forza k ed inversamente proporzionale alle masse degli atomi Essendo l’energia quantizzata, le transizioni saranno regolate da regole di selezione (∆ ∆n = ±1) e l’energia necessaria per la transizione (E vib ) sarà espressa dalla seguente relazione: ν osci 1 = 2π k µ 1 Evib = hν osci n + 2 Oscillatore armonico Le molecole reali sono meglio rappresentate da un modello di oscillatore anarmonico dove i livelli vibrazionali non sono ugualmente spaziati. Oscillatore anarmonico Armoniche superiori Overtones In un oscillatore armonico l’assorbimento di energia comporta il passaggio da un livello vibrazionale a quello immediatamente superiore (∆n= 1) In un oscillatore anarmonico ∆n può essere >1 anche se questo evento è poco probabile. Calcolo delle frequenze di assorbimento La legge di Hooke può essere applicata in maniera approssimata al calcolo delle frequenze assorbimento di molecole biatomiche (o di particolari legami) considerando come costanti di forza i seguenti valori: Legame singolo: k= 5 x 105 dyn/cm. Legame doppio: k= 10 x 105 dyn/cm Legame triplo: k= 15 x 105 dyn/cm. Ad es. per un legame C=C : ν = 1682 cm-1 (calcolata); 1650 cm-1 (exp.) Ad es.per un legame C-H: ν = 3032 cm-1 (calcolata); 2900 cm-1 (exp.) Diversi fattori influiscono sulla frequenza di assorbimento Masse degli atomi (⇑ massa, ⇓ ν) C-H C-C C-O C-Cl C-Br C-I 3000 1200 1100 750 600 500 cm-1 Forza di legame (⇑k forza, ⇑ ν) C≡C C=C C−C 2150 1650 1200 cm-1 Ibridazione (sp > sp2 > sp3) C-H (sp) C-H (sp2) C-H (sp3) 3300 3100 2900 cm-1 Tipi di vibrazione • Le vibrazioni possono essere divise in due categorie: – Stiramento (stretching): comportano il continuo cambiamento della distanza di legame tra gli atomi coinvolti – Deformazione o piegamento (bending): implicano la continua variazione dell’angolo di legame tra gruppi di atomi • Le vibrazioni di bending possono a loro volta essere di 4 tipi: – A forbice nel piano (scissoring) – Oscillazione nel piano (rocking) – Ondeggio fuori dal piano (wagging) – Torsione fuori dal piano (twisting) Tipi di vibrazioni: i modi di stretching Stretching simmetrico Stretching asimmetrico Tipi di vibrazioni: i modi di bending Rocking Scissoring Wagging Twisting Example of vibrational modes -CH2- . . . Stretching modes -CH2- asym. stretching νas(CH2) ~2926 cm-1 sym. stretching νs(CH2) Bending modes -CH2- scissoring δs(CH2) rocking ρ(CH2) ~2853 cm-1 ~1465 cm-1 ~725 cm-1 wagging ω(CH2) x twisting τ(CH2) 1350-1150 cm-1 Regola di selezione Assorbimento della radiazione se la variazione di stato vibrazionale è associata ad una variazione del momento dipolare (µ)della molecola r ∂µ r≠0 ∂r r ∂µ ε ≈ r ∂r Intensità banda IR aumenta con la variazione del momento dipolare Modi normali di vibrazione Quante modi normali sono possibili in una molecola? Modi normali: posizioni relative degli atomi cambiano ma la posizione e l’orientazione “medie” della molecola rimangono fisse Una molecola ha un numero di gradi di libertà uguale al grado di libertà totale di tutti i suoi atomi: molecola con N atomi : 3N gradi di libertà Esempio della CO2 Perché non si osservano 3N-5/3N-6 bande negli spettri ? Bande aggiuntive -Bande overtones (armoniche): transizione simultanea di due o più livelli del medesimo modo normale (2ν, 3ν) -bande di combinazione: transizione simultanea di due livelli vibrazionali corrispondenti a due diversi modi normali (ν1 +ν2) Bande assenti -nessuna variazione del momento dipolare -energia di vibrazioni diversi sono identiche (modi degeneri) -assorbimento troppo basso -ν fuori del range spettrale della misura Part II Aspetti Sperimentali Tipi di spettrometri To separate IR light, a grating is used. Grating Detector Slit In order to measure an IR spectrum, the dispersion Spectrometer takes several minutes. Also the detector receives only a few % of the energy of original light source. Sample Light source Fixed CCM To select the specified IR light, A slit is used. An interferogram is first made by the interferometer using IR light. Detector B.S. Sample Moving CCM IR Light source Spettrometro dispersivo The interferogram is calculated and transformed into a spectrum using a Fourier Transform (FT). Spettromettro a trasformata di Fourier In order to measure an IR spectrum, FTIR takes only a few seconds. Moreover, the detector receives up to 50% of the energy of original light source. (much larger than the dispersion spectrometer.) FTIR seminar FT optical system diagram Light source He-Ne gas laser (ceramic) Beam splitter Movable mirror Sample chamber (DLATGS) Fixed mirror Interferometer Detector Interferometria La ricombinazione tra i due raggi genera un’interferenza che dipende dal ritardo ottico tra i due raggi. Intensità misurata con il detector Trasformata di Fourier Tipico Interferogramma di una sorgente IR APPLICANDO LA TRASFORMATA DI FOURIER DELL’INTERFEROGRAMMA, SI PUO RICAVARE LO SPETTRO, B(ν) DELLA RADIAZIONE TRASMESSA DA UN CAMPIONE POSTO DAVANTI AL DETECTOR Interferometro Esistono diversi tipi di interferometro: -con ottiche piane (interferometro di Michelson) -con ottiche a corner cube Michelson Disallineamento di 1 µm Variazione interferogramma >10% Corner Cube Allineamento perenne Attrito : Se c’è attrito nella movimentazione dello specchio mobile, vita media finita che dipende dell’uso della macchina Rimozione attrito: Cuscino d’aria / Levitazione magnetica Il laser L’interferogramma viene generato mediante il campionamento fatto con un laser He-Ne (λ = 632 nm, 15800 cm-1) Campionamento delle intensità dell’interferogramma in un intervallo temporale, ∆t, sincronizzato con l’intensità della radiazione laser modulata dallo specchio mobile dell’interferometro. La frequenza di campionamento è data da: V: velocità spostamento dello specchio mobile mod f laser = 2Vvlaser DETECTOR VELOCE PERMETTE UNA VELOCITA DI CAMPIONAMENTO ELEVATA POSSIBILE REGISTRARE NUMEROSI INTERFEROGRAMMA AL SECONDO (ogni 20-30 µs) Componenti -SORGENTE Radiazione si produce passando una corrente elettrica attraverso un elemento riscaldante SORGENTI PIU COMUNI PER IL MIR -Sorgente di Nernst (Misc. Zirconio, Ittrio, Torio) : 2500-500 cm-1 -SPIRA Ni-Cromo : 13000-500 cm-1 -GLOBAR (Carburo di Silicio) : 8000-250 cm-1 -SEPARATORE DI RAGGIO (BEAM SPLITTER) Materiale semi-riflettivo depositato su un materiale trasparente Mylar Silicon su CaF2 Ge su KBr Quarzo Intervallo di applicazione (cm-1) 600-150 50000-1400 5000-200 15000-3500 Componenti -DETECTOR -Termici: funzionano a Troom con largo intervallo di frequenza (SVANTAGGI : LENTI, SCARSA SENSITIVITA) - Piroelettrici (cristalli piroelettrici sotili): -Deuterated Triglycine Sulfate (DTGS) -Lithium Tantalate (LITA) - Fotoconductivi (Alta sensitività, veloce) -Mercuri Cadmium Telluride (MCT) -Indium antimonite (InSb) Intervallo di Frequenza applicazione (cm-1) campionamento (kHz) DTGS MCT InSb InGaAs 5000-200 10000-600 10000-1350 12000-4000 6-10 40-150 1-4 InSb Sensibilità Spettrale Detectors MCT DTGS Generation of a FT-IR spectrum Step 1: scanning air (empty beam) R(ν ν) ratio between the two single-beam spectra flat baseline Step 2: scanning the sample S(ν ν) % Transmittance S (ν ) T (ν ) = 100 × R(ν ) Risoluzione spettroscopica Assorbanza (u.a.) Capacità di risolvere due righe spettrali R = 2 cm-1 R = 4 cm-1 R = 10 cm-1 R = 20 cm-1 1400 1300 1200 1100 1000 900 800 Numero d'onda (cm ) -1 Criterio di Rayleigh Per risolvere due righe spettrali separate da una distanza d (cm-1) si deve misurare un’interferogramma almeno su un camino L = 1/d La risoluzione massima di uno spettrometro è uguale al camino ottico massimo consentito dall’interferometro Parametri che influenzano la risoluzione FUNZIONE DI TRONCAMENTO Consente di effettuare un troncamento dell’interferogramma () +∆ B ' ν = ∫ I (δ ) D(δ ) cos(2πδν )dδ () () () −∆ B' ν = D ν ⋅ B ν Perdità di risoluzione Scelta funzione critica per: -bande strette (gas) -band fitting Spettri in fase gas Boxcar assicura la migliore risoluzione Misure quantitative con films polimerici Blackman-Harris Rapporto segnale/rumore (SNR) Definito come l’inverso del livello del rumore sulla linea 100% prodotta dividendo due single beam consecutivi misurati nelle stesse identiche condizioni Da cosa dipende? •Componenti ottiche (sorgente, beamsplitter, detector) •Parametri di misura (risoluzione, tempo di misura, velocità specchio,…) SNR ≈ t = tsin gle NS ≈ NS Per aumentare di un ordine di grandezza il SNR rispetto ad una scansione bisogna mediare 100 scan SNR ≈ ∆ν SNR aumenta con la risoluzione Advantages of FTIR spectroscopy • Speed All the frequencies are recorded simultaneously; a complete spectrum can be measured in less than a second. • Sensitivity In the interferometer, the radiation power transmitted on to the detector is very high which results in high sensitivity. • Internal calibration FTIR spectrometers employ a HeNe laser as an internal wavelength calibration standard, no need to be calibrated by the user. • Multicomponent capability Since the whole infrared spectrum is measured continuously, all infrared active components can be identified and their concentrations determined. Ideal to examine broad spectral regions and weak absorptions with high resolution Principal types of sampling holding TRANSMISSION SAMPLING (traditional method) Ir Io SAMPLE I0 REFLECTION SAMPLING Detector I It Advantages: -highest S/R ratio -No distorsion of the spectra -Quantitative analysis straightforward - Spectral librairies with trans. Meas. Disadvantages: -Need for thin samples (liquid<2µm, film: 10-100µm) Advantages: - No need for thin samples Disadvantages: -low S/R ratio -Distorsion of the spectra -Complex quantitative analysis Transmission sample handling LIQUIDS GASES High Sensitivity (Multipass) Sample Cell L = 60 … 980 cm Demountable liquid cell with teflon spacers L = 0.015 to 1 mm Single pass cell L = 1, 4, or 10 cm Scelta delle celle Il campione deve essere contenuto in celle che non devono assorbire le radiazioni IR nelle zone di interesse Cut-off NaCl : 625 cm-1 KBr : 400 cm-1 CsI : 200 cm-1 SOLUBILITA CELLA !!!! KBr : solubile in H2O AgBr : insolubile in H2O Solid samples (powder) i)Grind and mix with Nujol (hydrocarbon oil) to form mull ii)Make semi-transparent pellet with KBr (1-3 mg sample in 250-300 mg KBr). • Halide salts get transparent when pressured ARTEFACTS (a) increase of light loss from reflection and scattering by large particles 55µ 40µ 23µ 15µ 5µ 2µ CaCO3 in KBr, showing the mean diameters of the absorbing particles (b)matching of sample and medium RI to prevent scattering. e.g. PVC (nD = 1.548) dispersed in KBr (nD = 1.56), KCl and KI. KBr KCl KI (c) Fattori chimici o fisici (es. Reazione con Halide) Spectra of benzoic acid in alkali halide discs. Spettro IR del Nujol 721 1462 2925 2855 1377 Possible problems with spectra Asymmetric, sloping bands. Badly ground. Sample does not cover beam. Also for air bubble in liquid cell; polymer film with hole or crack Sample (mull) too thick Reflectance sampling handling Diffuse Reflectance (DRIFT) Ir out Transmission-Reflection (absorption-reflection) IR in IR in Attenuated Total Reflection (ATR) IR in IR out sample IR out Diffuse reflectance (DRIFT) Diffuse reflection by a powder Principle of the method Measurement of the energy that has been partially absorbed, transmitted and partially scattered by a surface with no defined angle of reflection Optical arrangement for DRIFT Applications -strongly absorbing samples -small, irregular samples, powders -surface analysis of hard objects using abrasive paper Advantages -minimal sample preparation -sample not destroyed Attenuated total reflection (ATR) IR in IR out sample Principle of the method when radiation strikes an interface with a medium of lower RI total reflection can occur for an angle > θc (sin θc = nsam / ncrys) Radiation actually BUT penetrates sample and is partially absorbed Applications -strongly absorbing samples (films, pastes, gel, liquid) -Surface analysis Advantages -Minimal sample preparation -Possible to analyzed multicoated samples Choice of ATR crystal Crystal material defines the RI, the IR range, and the chemical properties of the crystal crystal material IR range cm-1 RI at 2000 cm-1 general properties KRS-5 17000-400 2.38 Soft Soluble in bases Slightly soluble in water ZnSe 20000-650 2.43 Insoluble in water and organic solvents 4.01 Hard & brittle Insoluble in water 2.4 Very Hard Chemically inert Ge 5500-675 Diamond 4500-2500 &1680-30 Distance where the intensity of the evanescent wave decays to 1/e dP = 1 2πνncris sin 2 θ − (ncamp ncris ) 2 (dp=0.5-2 µm) dp is “equivalent” to the sample thickness for transmission sampling Factors affecting ATR spectra •Angle of incidence (θ):dp (and thus A) decreases with increasing θ (helpul for depth profile near the surface) •RI of the crystal (ncris): dp (and thus A) decreases with increasing RI •Wavenumber (ν): dp (and thus A) decreases with increasing ν (ATR spectrum differs from the analogous transmission spectrum) • Efficiency of sample contact: dp (and thus A) decreases with decreases the sample contact ATR-FTIR of China Clay Filled Polyester Film 30o Effetto del angolo di incidenza Variazione angolo 30°- 60° diminuzione d di un fattore ~10. Can depth profile by changing θ. Smaller angle, deeper penetration 45o 60o Celle ATR Transmission-Reflection IR in Principle of the method By placing the sample on a reflective surface, IR radiation travels through the sample, is reflected and passes a second time the sample. Applications -coating on reflective surface (0.5-20 µm thickness range) Advantages -spectrum similar to transmission one -double pathlength IR out Coating of soft drink can IR microspectrometry • • • • Introduced in 1980s IR transmission or reflection spectra Sample dimensions in 10 -500 µm Instrument – Ordinary optical microsocpe – FT-IR with small IR beam size – LN2 MCT (mercury/cadmium/telluride) PC Advantages -Possible to analyze small samples -IR imaging FTIR microscope Transmission measurement Cross-section of food packaging multilayer film cut to 10 µm thickness by microtome. (1), (3) are PE and PP H2O barriers (2) Is ethylene vinyl alcohol copolymer, O2 barrier (1) 60 µm (3) 40 µm (1) (2) (3) Microscopic IR spectra of layers (2) 8 µm Utilizzo di fibre ottiche Fibre ottiche possono essere usate per misure in riflettanza. Con l’impiego di sonde è possibile irraggiare il campione e raccogliere la radiazione diffusa; entrambe le radiazioni possono viaggiare su fibre Vantagggi: -Possibilità di effettuare analisi in situ, totalmente non distruttive e senza toccare il campione; -Non ci sono vincoli dovuti alla forma del campione Fibre ottiche :vetri a calcogenuro (As-Se-Te) o calcoalogenuro. nessuna delle fibre attualmente disponibili è in grado di coprire l’intero range del medio infrarosso; (massimo 4000-900 cm-1) Sonde per riflettanza diffusa La sonda per riflettanza diffusa può essere costituita da una fibra ottica per l’irraggiamento e una per la raccolta della radiazione diffusa Sonde ATR Nessuna limitazione al range spettrale analizzabile se non in funzione del cristallo ATR Caratterizzazione di film sottili mediante spettroscopia IRRAS (Infra Red Reflection Absorption Spectroscopy) Tecnica di campionamento deve essere adattata allo spessore dei campioni spessore = 10 - 200 µm Trasmissione spessore > 200 µm Riflessione Totale Attenuata (ATR), DRIFT spessore < 1 µm IRRAS (Infra Red Reflection Absorption Spectroscopy) Film sPS Spessore: 100 nm Trasmissione IRRAS Principio di funzionamento della tecnica IRRAS Film da analizzare deve essere depositato su una superficie metallica riflettente Ep Il piano di riflessione (o d’incidenza) è individuato dal raggio incidente e dalla normale alla superficie riflettente Es due stati di polarizzazione sono definiti: •Es polarizzato ortogonalmente al piano di riflessione (parallelamente al piano del film) •Ep polarizzato parallelamente al piano di riflessione (perpendicolarmente al piano del film) cambiamento di fase per la luce riflessa dalla superficie metallica al variare dell’angolo d’incidenza Es Es Ep Ep Cambiamento di fase per i due stati di polarizzazione Es ed Ep: a) stato di polarizzazione Es incidenza obliqua alla superficie b) stato di polarizzazione Es incidenza normale alla superficie c) stato di polarizzazione Ep incidenza normale alla superficie d) stato di polarizzazione Ep incidenza obliqua alla superficie •Stato di polarizzazione Es, i due vettori campo elettrico (incidente e riflesso) si annullano nel punto di contatto l’assorbimento della componente s è nulla. •Stato di polarizzazione Ep, nel caso d’incidenza obliqua, i due vettori campo elettrico (incidente e riflesso) non si annullano nel punto di contatto e la loro risultante è perpendicolare alla superficie metallica Coefficienti di assorbimento calcolati per gli stati di polarizzazione Es ed Ep in funzione dell’angolo d’incidenza (θ) (strato di acetone di 0.1 nm) •per lo stato di polarizzazione Ep il massimo assorbimento si ha per angoli d’incidenza prossimi a 90° •per lo stato di polarizzazione Es assorbimento costante e molto basso In sintesi: 1)L’angolo di incidenza ottimale per effettuare una misura IRRAS deve essere vicino all’angolo radente (c.a. 88°) 2)Solo lo stato di polarizzazione Ep della radiazione incidente è assorbito da un film sottile o da molecole depositate su una superficie metallica. 3)La variazione del momento dipolare di transizione della molecola depositata o del film sottile deve avere una componente perpendicolare alla superficie per dare luogo ad un assorbimento della radiazione. COMME EFFETUARE LA MISURA? Metodo classico: 1. Misura bianco con solo supporto 2. Misura con film Problema: L’intensità assorbita dal film è molto bassa SNR basso Alternativa: PM-IRRAS: Polarization Modulation-Infra Red Reflection Absorption Spectroscopy •sorgente IR •filtro ottico •polarizzatore lineare •modulatore fotoelastico •portacampione •lente convergente •Detector (MCT) Il modulatore fotoelastico permette di di modulare lo stato di polarizzazione della radiazione IR (stati di polarizzazione Es e Ep) L’intensità riflessa dal campione è misurata con il detector per i due stati di polarizzazione P e S (misura simultanea): Segnale α Ip+Is Segnale α Ip-Is •Ip e Is sono attenuati nello stesso modo dalle molecole non orientate (H2O, CO2) •Soltanto Ip è attenuata dal film sottile R= (Ip-Is)/(Ip+Is) permette di ottenere unicamente l’assorbimento dovuto al film Confronto spettri IRRAS/PM-IRRAS IRRAS PM-IRRAS Polilisana (bio-polimero) Near-infrared spectrometry NIR shows some similarities to UV-visible spectrophotometry and some to mid-IR spectrometry. Indeed the spectrophotometers used in this region are often combined UV-visible-NIR ones. The majority of the absorption bands observed are due to overtones (or combination) of fundamental bands that occur in the region 3 to 6 µm, usually hydrogen-stretching vibrations. NIR is most widely used for quantitative organic functional-group analysis. The NIR region has also been used for qualitative analyses and studies of hydrogen bonding, solute-solvent interactions, organometallic compounds, and inorganic compounds. Bande overtones (armoniche): transizione simultanea di due o più livelli del medesimo modo normale (2ν, 3ν) Bande di combinazione: transizione simultanea di due livelli vibrazionali corrispondenti a due diversi modi normali (ν1 +ν2) 14280 cm-1 4000 cm-1 Far-infrared spectrometry The far-IR region is used to study low frequency vibrations, like those between metals and ligands (for both inorganic and organometallic chemistry). The fundamental vibrations of many organometallic and inorganic molecules fall in this region due to the heavy atoms and weak bonds in these molecules. Example: Metal halides have stretching and bending vibrations in the 650-100 cm-1 range. Common applications of IR spectroscopy -Identification of compounds by matching spectrum of unknown compound with reference spectrum (fingerprinting) -Identification of functional groups in unknown substances -Identification of reaction components and kinetic studies of reactions -Identification of molecular orientation in polymer films - Detection of molecular impurities or additives present in amounts of 1% and in some cases as low as 0.01% - Identification of polymers, plastics, and resins - Analysis of formulations such as insecticides and copolymers But also -Measurement of branching (HDPE, LDPE) -Polymorphism in polymer (chain conformation, chain packing) - Polymer blends (compatibility, degree of mixing) -… Spettroscopia Raman Effetto Raman (1928): diffusione anelastica della radiazione elettromagnetica dovuta all’interazione con i moti rotazionali e vibrazionali della molecola Effetto Raman luce di energia minore di quella richiesta per promuovere una molecola in uno stato elettronico eccitato viene assorbita dalla molecola si crea uno stato eccitato virtuale. maggior parte della luce viene riemessa in tutte le direzioni alla stessa energia: Rayleigh scattering L’energia di una piccola porzione di luce riemessa differisce dalla radiazione incidente di intervalli di energie che corrispondono ad alcuni modi vibrazionali. energie più alte (ν νo + hν ν) : la differenza di energia deriva da una perdita di energia vibrazionale della molecola (righe anti-Stokes) energie minori (ν νo - hν ν) si è verificata una promozione vibrazionale (righe Stokes). •IRaman << IRayleigh (10-8) •Ianti-Stokes <IStokes stokes antistokes Spettro Raman del CCl4 Rayleigh line (elastic scattering) ν0 = 20492 cm-1 λ0 = 488.0 nm Stokes lines (inelastic scattering) Observed in “typical” Raman experiments Anti-Stokes lines (inelastic scattering) 459 314 218 -218 -314 -459 400 200 0 -200 -400 Raman shift cm-1 ν0 = (ν νs - µ0) Figure is redrawn from D. P. Strommen and K. Nakamoto, Amer. Lab., 1981, 43 (10), 72. Ligne Stokes e anti-stokes sono simmetriche rispetto alla linea Rayleigh e la differenza in energia corrisponde alla variazione del livello vibrazionale iniziale Spettro Raman Intensità di emissione luminosa v.s. SHIFT RAMAN ∆ν ∆ν: differenza tra radiazione diffusa emessa e radiazione sorgente ∆ν cade nel medio IR ed è indipendente della lunghezza d’onda della radiazione incidente NB: Per lo spettro si prendono le linee Stokes (indicate in cm-1 positivi) POSIZIONE E INTENSITA delle bande sono legate ai gruppi funzionali e ai loro modi di vibrazione in maniera analoga alla spettroscopia IR. FT-IR Flufenamic acid Aldrich as recd 0.30 0.25 Abs 0.20 0.15 0.10 GRANDI DIFFERENZE TRA LE VIBRAZIONI ATTIVE ALL’IR E AL RAMAN 0.05 FT-Raman Flufenamic acid Aldrich as recd 60 50 Int 40 30 20 10 0 3500 3000 2500 2000 Raman shift (cm-1) 1500 1000 500 Vibrazioni di CO2 Stiramento simmetrico Stiramento asimmetrico Raman Attivo IR Attivo + Piegamenti IR Attivi - + Vibrazione IR inattiva Polarizzabilità Nessun momento di dipolo Raman attiva Vibrazione IR attiva Momento di dipolo oscillante Livelli energetici di CO2 Differenze di energia, cm-1 20,000 Raman 2350 1335 667 Zero IR IR % Trasmissione Spettri IR e Raman di CO2 Spettro IR Intensità diffusa 4000 2000 0 2000 Numeri d’onda (cm-1) 0 Spettro Raman 4000 Un modo vibrazionale può essere IR attivo, Raman attivo od entrambi Regola di mutua esclusione Molecola con un centro di simmetria nessun modo può essere sia IR che Raman attivo Esecuzione di una misura Raman Per ottenere lo spettro Raman di un campione si utilizza una sorgente monocromatica con lunghezza d’onda nel vicino UV, nel visibile o nel vicino infrarosso (NIR). Attualmente sono impiegate sorgenti LASER; alcuni esempi sono i seguenti: • Laser UV, 244 o 325 nm • Laser ad Argon, 488.0 nm o 514.5 nm • Laser a He-Ne, 632.8 nm • Laser a rubino, 694.3 nm • Laser Nd-YAG (ittrio–alluminio–granato drogato con neodimio), 1064 nm La radiazione laser è focalizzata sul campione; le radiazioni diffuse dalla superficie sono raccolte, rivelate dal detector e mostrate sotto forma di spettro Utilizzo di laser diversi SPOSTAMENTO RAMAN, per un dato legame ed un determinato modo di vibrazione, è indipendente dalla λ di eccitazione (i.e. sorgente laser) spettri Raman espressi in unità di Raman shifts sono in teoria identici con qualunque laser vengano prodotti e sono confrontabili anche se prodotti con strumenti Raman diversi Tuttavia utilizzo di laser diversi su uno stesso campione puo generare segnali Raman nelle stesse posizioni ma con intensità diverse Con i pigmenti, si verifica sperimentalmente che il laser verde (514.5 nm) è più adatto nell’ottenere spettri di pigmenti blu e verdi, mentre il laser rosso (632.8 nm) è più adatto per pigmenti rossi e gialli. In certi casi alcuni pigmenti non forniscono alcun spettro se non sono irraggiati con il laser più opportuno Esempio effetto laser su un dye Scelta del laser La frequenza del laser impiegato, e quindi l’energia irraggiata sul campione, influenzano in maniera decisiva lo spettro Raman che si può ottenere. È necessario tenere conto dei seguenti aspetti: • l’intensità di emissione Raman è proporzionale alla quarta potenza della frequenza della sorgente: in altre parole, l’emissione ottenibile con un laser UV a 244 nm (40984 cm-1) in teoria è enormemente più intensa di quella ottenibile con un laser NIR a 1064 nm (9399 cm-1) • l’energia in gioco con laser a più alta frequenza (UV, visibile) è in grado di attivare nel campione transizioni elettroniche non desiderate che possono generare fenomeni di fluorescenza e produrre spettri di difficile lettura; questi fenomeni sono meno evidenti con laser meno energetici come il NIR • un altro inconveniente dei laser ad alta frequenza è il danno che possono causare ai campioni durante l’irraggiamento, causando fotodecomposizione e quindi emissione di spettri Raman anomali CONFRONTO IR - RAMAN le due tecniche sono più complementari che competitive IR RAMAN ASSORBIMENTO DIFFUSIONE ANELASTICA VARIAZIONE DEL MOMENTO DI DIPOLO (forte: legami ionici come O-H, N-H) VARIAZIONE DELLA POLARIZZABILITA’ (forte: legami covalenti come C=C, C-S) PREPARAZIONE DEL CAMPIONE SPESSORE OTTIMALE (TRASMISSIONE) o CONTATTO CON IL CAMPIONE (ATR) NESSUN CONTATTO, NESSUNA DISTRUZIONE DEL MATERIALE, SEMPLICE PREPARAZIONE; H2O COME SOLVENTE O VETRO COME CELLA NON DISTURBANO LA MISURA VANTAGGI SPETTRI PIU RICCHI lINNEE DELLO SPETTRO PIU STRETTE (PIU FACILI DA IDENTIFICARE) PROBLEMI FORTE ASSORBIMENTO DI VETRO, ACQUA, CO2 FLUORESCENZA MATERIALI PRINCIPALMENTE COMPOSTI ORGANICI PRESSOCHE’ SENZA LIMITI INTERVALLO DI FREQUENZE 4000 - 400 cm-1 (MIR) 4000 - 50 cm-1 EFFETTO FISICO PER LE SOSTANZE INORGANICHE, i segnali Raman relativi ai legami metallo-legante sono generalmente nel range 100-700 cm-1, (regione spettrale di difficile utilizzo nell’IR)