C40
IAE spa
via F.Filzi 53
20032 CORMANO MI
tel. +39 02 66303255
fax +39 02 6152544
e-mail [email protected]
Rotating anode X-ray tube unit
for general diagnostic systems
Application
Rotating anode X-ray
tube unit specifically
designed for 80 mm
anode diameter inserts,
for radiographic and
fluoro procedures.
Characteristics
•
•
•
•
•
Lead lined aluminium body.
Filled under vacuum with insulating oil, specially processed.
Two devices are present for thermal safety:
1. a bimetallic thermal switch, internally assembled, series connected with stator common cable
2. a pressure switch
Storage and shipment temperature range -10°C ÷ +80°C
No user maintenance requested
Inserts for reloading
•
•
X39
X42
Specifications
Overall length
489 mm
Maximum diameter
152 mm
C40 Rev 0 - 21.02.2013
Tube assembly net weight
16 kg
Nominal X-ray tube assembly voltage
125 kV
Maximum tube assembly heat content
900 kJ
Maximum continuous heat dissipation
180 W
Minimum tube assembly inherent filtration
Maximum leakage radiation at 1 m from focal spots
1.0 mm Al / 75 kV
0.44 mGy/h
Dimensions
489
156
170
Ø140
41
81
4-M6 8 Deep
Asse di riferimento
Reference axis
Axe de rèfèrence
76.2
Ø70
F
81
52
X39 - X42
121.6
Asse di riferimento
Reference
Axe de rèfèrence
Tube assembly heating and cooling curves
Energia Accumulata - Stored Energy - Energie
Emmagasinée (kJ)
900
800
300 W
700
200 W
600
500
150 W
400
100 W
300
50 W
200
100
0
0
20
40
60
80
100
120
140
160
180
200
220
240
Tempo Time - Temps (min)
Should you have questions about the compatibility of a replacement tube unit call our customer service: +390266303255.
© copyright iae spa
Data may vary without notice
IAE spa
via F. Filzi 53
20032 Cormano MI - Italy
tel. +39 02 66303255
fax +39 02 6152544
e-mail [email protected]