C40 IAE spa via F.Filzi 53 20032 CORMANO MI tel. +39 02 66303255 fax +39 02 6152544 e-mail [email protected] Rotating anode X-ray tube unit for general diagnostic systems Application Rotating anode X-ray tube unit specifically designed for 80 mm anode diameter inserts, for radiographic and fluoro procedures. Characteristics • • • • • Lead lined aluminium body. Filled under vacuum with insulating oil, specially processed. Two devices are present for thermal safety: 1. a bimetallic thermal switch, internally assembled, series connected with stator common cable 2. a pressure switch Storage and shipment temperature range -10°C ÷ +80°C No user maintenance requested Inserts for reloading • • X39 X42 Specifications Overall length 489 mm Maximum diameter 152 mm C40 Rev 0 - 21.02.2013 Tube assembly net weight 16 kg Nominal X-ray tube assembly voltage 125 kV Maximum tube assembly heat content 900 kJ Maximum continuous heat dissipation 180 W Minimum tube assembly inherent filtration Maximum leakage radiation at 1 m from focal spots 1.0 mm Al / 75 kV 0.44 mGy/h Dimensions 489 156 170 Ø140 41 81 4-M6 8 Deep Asse di riferimento Reference axis Axe de rèfèrence 76.2 Ø70 F 81 52 X39 - X42 121.6 Asse di riferimento Reference Axe de rèfèrence Tube assembly heating and cooling curves Energia Accumulata - Stored Energy - Energie Emmagasinée (kJ) 900 800 300 W 700 200 W 600 500 150 W 400 100 W 300 50 W 200 100 0 0 20 40 60 80 100 120 140 160 180 200 220 240 Tempo Time - Temps (min) Should you have questions about the compatibility of a replacement tube unit call our customer service: +390266303255. © copyright iae spa Data may vary without notice IAE spa via F. Filzi 53 20032 Cormano MI - Italy tel. +39 02 66303255 fax +39 02 6152544 e-mail [email protected]