1. GAS S.Bianco Colafranceschi Benussi Piccolo, … 2. SENSORS4UPGRADE S.Bianco S.Buontempo Caponero Breglio…, … Stefano Bianco per CMS Frascati s.bianco (frascati) CMS RPC Italia 12 maggio 1 GAS Gas review al cern sull’incident della rottura del flussimetro SF6 (0.30% --> 0.39%) piuttosto produttiva: https://indico.cern.ch/conferenceDisplay.py?confId=137611 1. e minute di Pigi s.bianco (frascati) CMS RPC Italia 12 maggio 2 SF6 Mass Flow Controller – follow-up 15/4 possible problem concerning the RPC gas mixture composition reported the SF6 mass flow controller (MFC). MFC changed (19:30) and to start preparing the gas chromatograph (GC). 17/4 17:00 after discussion with RPC group, I put back the old MFC Analyses results: the old MFC is delivering about 30% more SF6 with respect the new. SF6 C2H2F4 iC4H10 s.bianco (frascati) CMS RPC Italia 12 maggio 3 SF6 Mass Flow Controller 18/4 20:00 New MFC again in use 19/4 New analysis confirmed the 34% difference between the two MFCs (i.e. 0.4 % instead of 0.3 % of SF6 in the mixture). 20/4 21:50 SF6 concentration in the return mixture from the detector: 0.325% old MFC verified on the our Bronkhorst calibration stand: 35% difference in the output flow confirmed. 21/4 12:00 SF6 concentration in the return mixture from the detector: 0.323% 21:00 SF6 concentration in the return mixture from the detector: 0.322% 25/4 21:00 SF6 concentration in the return mixture from the detector: 0.314% 3/5 Old MFC reinstalled after recalibration 21:00 SF6 concentration in the return mixture from the detector: 0.312% Regular analysis (1-2 per week Beatrice) will be performed until when the CMS GC will be fully operational s.bianco (frascati) CMS RPC Italia 12 maggio 4 GGM on wrong mix problem The GGM HV scan confirms shift of WP due to wrong mix In progress analysis of charge ratios timeline to spot (wrongMix->goodMix) transition which should have occurred delayed in clean and (after/before) gas mixtures. A small transient effect possible. s.bianco (frascati) CMS RPC Italia 12 maggio 5 PRELIMINARY GGM down :-(((( +-10% after down 2,500 charge ratio 2,000 1,500 1,000 0,500 0,000 1301000000 -0,500 1301500000 1302000000 1303000000 1302500000 1303500000 1304000000 1304500000 time APRIL 2011 s.bianco (frascati) CMS RPC Italia 12 maggio HV scans 6 after down 0.30%-->0.39% SF6 1,500 1,300 charge ratio 1,100 0,900 0,700 0,500 0,300 1302840000 1302940000 1303040000 1303140000 1303240000 1303340000 1303440000 1303540000 time s.bianco (frascati) CMS RPC Italia 12 maggio 7 after down 14 12 10 adc 8 6 4 2 0 1302840000 1302940000 1303040000 1303140000 1303240000 1303340000 1303440000 1303540000 1303340000 1303440000 1303540000 time clean down charge ratio ADC 20 15 10 5 0 1302840000 1302940000 1303040000 1303140000 1303240000 time s.bianco (frascati) CMS RPC Italia 12 maggio 8 after down 20 adc 15 10 5 0 1301000000 1301500000 1302000000 1302500000 1303000000 1303500000 1304000000 1304500000 1303500000 1304000000 1304500000 time charge ratio ADC clean down 20 15 10 5 0 1301000000 1301500000 1302000000 1302500000 1303000000 time s.bianco (frascati) CMS RPC Italia 12 maggio 9 CONCLUSIONS 0/3 1. 2. Il gas group effettuera’ misure di GC con il sistema CMS in USC 1/2 volte a settimana (Beatrice) Il GC INFN verra’ installato in SGX5 e operato da RPC s.bianco (frascati) CMS RPC Italia 12 maggio 10 CONCLUSIONS 1/3 • GGM : – Stable operation, performance up to specs. • INFN GC : – Install and operate in SGX5 gas room – Issues: automatize & integrate DCS/PVSS – runs : ~2 days frequency s.bianco (frascati) CMS RPC Italia 12 maggio 11 CONCLUSIONS 2/3 •The GGM is not designed for detecting overall gas mixture changes •Studies in progress to verify the feasibility of detecting gas mixture changes via the transient in the charge ratios. Small effects. • We propose to widen the scope and performance of GGM in order to provide an additional mean of monitoring for the RPC gas mixture • GGM+ : – HV feedback. Change HV supply to gaps as a function of p and T. Monitor charge and ratios • • • • PVSS changes including PVSS integration… 0.5 month Unix data analysis… 1.5 month Commissioning and debugging…….. 0.75 month Value added: Gain experience for a CMS RPC feedback – HV scans. Every 10 days • • • • PVSS changes …0.5 month Auto scan tool C++/ROOT … 2.0 months XDAQ changes… 0.5 month Commissioning and debugging 0.5 month • Need 6.5 months support for a Frascati engineer (R.Aurilio or C.Vendittozzi). s.bianco (frascati) CMS RPC Italia 12 maggio 12 CONCLUSIONS 3/3 1. Abbiamo trovato e capito alcuni GGM analysis bugs 2. Il transiente si vede 3. A Frascati stiamo facendo HVSCAN con twin GGM • SF6 = 0.30/0.39/0.33/0.30/0.25 • Circa altri 10 giorni 4. A CERN abbiamo iniziato serie di HV scan con GGM per studiare sistematica e alternative al fit con la sigmoide s.bianco (frascati) CMS RPC Italia 12 maggio 13 TASKS & RICHIESTE 1. Install INFN GC and operate in SGX5 • 1 mese … Beatrice (PV) oppure Vendittozzi (Frascati) 2. GGM+ • 6.5 mesi … Vendittozzi (Frascati) o Aurilio (Frascati) o ??? Altra risorsa RPC (Pakistan/India etc) s.bianco (frascati) CMS RPC Italia 12 maggio 14 POSSIBILE RIORGANIZZAZIONE GAS • GAS () 1. 2. 3. 4. Gas quality …. GC () Gas calibration ….GGM (Frascati) Liason to gas group (Benussi) Detector (leak tests + ???) () s.bianco (frascati) CMS RPC Italia 12 maggio 15 SENSORS4RPC Frascati + Napoli • Installare sensori ottici FBG per temperatura camere endcaps nuove +esistenti • 2013 ma vogliamo iniziare a comprare, certificare, saldare sensori al polipo ASAP • Oggi distribuito primo draft di specifiche tecniche gara • Next: riunioni frascati/napoli • Lab Frascati pronto per test accettazione, Napoli anche disponibile s.bianco (frascati) CMS RPC Italia 12 maggio 16 SPARE SLIDES s.bianco (frascati) CMS RPC Italia 12 maggio 17 RPC GAS MONITORING 1. Gas Gain Monitoring (GGM) • Monitoring of working point for changes due to differences among clean, afterpurifiers, before-purifiers gas mixtures. Monitoring of charge ratios (afterPurifiers/clean, beforePurifiers/clean) allows cancellation of p,T,H effects 2. Gas Quality Monitoring • Monitoring of chemical composition of gas mixtures s.bianco (frascati) CMS RPC Italia 12 maggio 18 RPC GAS SYSTEM AND MONITORING Gas quality monitoring Purifiers GC+others (half wheel pickup points) GC FpH other Mixer Gas gain monitoring Humidifier RPCs #1 Fresh gas RPCs #2 Before purifiers CMS UXC55 USC55 SGX5 s.bianco (frascati) CMS RPC Italia 12 maggio RPCs #3 After purifiers DAQ Performance detector monitoring ENV. CONTROL gas parameters Ambient parameters 19 19 GGM • Eight single gap RPCs with charge readout in a cosmics telescope • Three sets of gaps: clean gas, beforepurifier gas, after-purifier gas. Cancellation of p,T by ratioing. • fully integrated in DCS/PVSS •In operation and shift monitoring since 2009 https://twiki.cern.ch/twiki/bin/view/CMS/RPComm s.bianco (frascati) CMS RPC Italia 12 maggio 20 GGM s.bianco (frascati) CMS RPC Italia 12 maggio 21 GGM s.bianco (frascati) CMS RPC Italia 12 maggio 22 Gas Gain Monitoring Performance Avalanche Streamer Anode charge in avalanchestreamer transition region monitors shifts in working point. SENSITIVITY ~ 20V 0% SF6 +-3%rms Coherent effects cancel out due to system redundancy. The two-gap ratio cancels out 0.3% SF6 s.bianco (frascati) CMS RPC Italia 12 maggio LONG-TERM ENVIRONMENTAL EFFECTS AT THE +-3% LEVEL 23 GGM on PVSS s.bianco (frascati) CMS RPC Italia 12 maggio 24 GGM TWIN IN FRASCATI CMS LAB T,H controlled hut Gaschromatograph DAQ system s.bianco (frascati) CMS RPC Italia 12 maggio +-1% +-0.5 C 25 GAS QUALITY MONITORING • Gaschromatograph (GC) with columns optimized for RPC gas mixture components • Purchased in ~2007 by the INFN group and used by CERN gas group so far • Nicely complement to the CMS GC presently operated by the CERN gas group in USC. s.bianco (frascati) CMS RPC Italia 12 maggio 26 STATUS OF GGM AND GAS QUALITY •GGM •Fully integrated in DCS/PVS •In operation and shift monitoring since 2009 https://twiki.cern.ch/twiki/bin/view/CMS/RPComm •Stable operation •Loss of efficiency in 1 gap, to be replaced • The GGM is performing up to specs • INFN Gaschromatograph •Not periodically used •No procedures for shifter use available yet s.bianco (frascati) CMS RPC Italia 12 maggio 27 OUTLOOK & GGM+ We propose actions to improve the monitoring of CMS RPC gas mixture: 1. Review of alarms on gas parameters (see G.Polese’s talk) 2. Integrate in the gas monitoring the INFN GC located at SGX5 gas room. 3. GGM+ • p,T correction via HV feedback • HV scans s.bianco (frascati) CMS RPC Italia 12 maggio 28 GGM HV feedback test with twin system in Frascati With p,T correction on HV Without p,T correction on HV s.bianco (frascati) CMS RPC Italia 12 maggio 29 TYPICAL GGM HV SCAN s.bianco (frascati) CMS RPC Italia 12 maggio 30 GGM preliminary results on wrong mix problem April 29th, 2011 - March 3rd, 2011 Good mix Wrong mix s.bianco (frascati) CMS RPC Italia 12 maggio April 29th, 2011 - January 2010 Good mix Good mix 31 RESOURCES 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. M. Abbrescia et al., Gas analysis and monitoring systems for the RPC detector of CMS at the LHC, [physics/0701014] LNF-06-34-P (Jan 2007) 9p. Presented by S.Bianco at the IEEE2006, S.Diego (USA). M.Abbrescia et al., The gas monitoring system for the resistive plate chamber detector of the CMS experiment at LHC. 2008. 4pp. 18th Hadron Collider Physics Symposium 2007 (HCP 2007) 20-26 May 2007, La Biodola, Isola d'Elba, Italy. Published in Nucl.Phys.Proc.Suppl.177-178:293-296,2008. CMS Collaboration, The CMS experiment at the CERN LHC, 2008 JINST 3 508004. L.Benussi et al., Sensitivity and environmental response of the CMS RPC gas gain monitoring system, JINST 4:P08006,2009,. [arXiv:0812.1710] L.Benussi et al., The CMS RPC gas gain monitoring system: An Overview and preliminary results. Nucl.Instrum.Meth.A602:805-808,2009,. [arXiv:0812.11]. L.Benussi et al, Sensitivity and environmental response of the CMS RPC gas gain monitoring system, LNF-08-27(P), JINST 4:P08006,2009. e-Print: arXiv:0812.1710. S.Bianco et al., Chemical study of the CMS RPC closed loop gas system, CMS note 2010 - 007 L.Benussi et al., Study of gas purifiers for the CMS RPC detector. [arXiv:1012.5511] (Dec 2010) L.Benussi et al., A New approach in modeling the response of RPC detectors. [arXiv:1012.5508] (Dec 2010) 13p. S.Colafranceschi et al., Operational experience of the gas gain monitoring system of the CMS RPC muon detectors. Nucl. Instrum. Meth.A617:146-147,2010,. L.Benussi et al., A new approach in modeling the response of RPC detector, presented by S.Colafranceschi at RPC10, Darmstadt, to appear on NIM (2011) G.Saviano (for the CMS RPC Collaboration) Results on chemical analysis in the CMS RPC closed loop, presented at RPC2010, Darmstadt (Germany). To appear on NIM (2011). https://twiki.cern.ch/twiki/bin/view/CMS/RPComm https://twiki.cern.ch/twiki/bin/view/CMS/MuonRpcPrivate s.bianco (frascati) CMS RPC Italia 12 maggio 32 SPARES s.bianco (frascati) CMS RPC Italia 12 maggio 33 GGM-related studies on purifiers • • • 1. 2. 3. • Study of RPC purifiers at the scaleddown closed loop system located in the ISR area Release of contaminants in gas is correlated to currents increase Results available at S.Bianco et al., Chemical study of the CMS RPC closed loop gas system, CMS note 2010 - 007 L.Benussi et al., Study of gas purifiers for the CMS RPC detector. [arXiv:1012.5511] (Dec 2010) G.Saviano (for the CMS RPC Collaboration) Results on chemical analysis in the CMS RPC closed loop, presented at RPC2010, Darmstadt (Germany). To appear on NIM (2011). Currently running, expect run ending in about 2 months, move test to GIF s.bianco (frascati) CMS RPC Italia 12 maggio 34