The installation of the new VIRGO Injection Bench Paolo La Penna for the VIRGO collaboration European Gravitational Observatory N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Recombined configuration: first half of 2004 3 km INPUT MODE CLEANER WE Rc urv= 3500 m R= 0.99995 (144 m) 5.6 m Nd:YAG = 1064 nm 20 W WI plane R= 0.88 IB Modulator single frequency 6.26 Mhz 3 km 6m 6.4 m BS NI plane R= 0.88 PR plane R= 0.925 NE Rc urv= 3500 R= 0.99995 OMC DETECTION SYSTEM N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Recycled interferometer: July 2004 3 km INPUT MODE CLEANER WE Rc urv= 3500 m R= 0.99995 (144 m) 5.6 m Nd:YAG = 1064 nm 20 W WI plane R= 0.88 IB Modulator single frequency 6.26 Mhz 3 km 6m 6.4 m BS NI plane R= 0.88 PR plane R= 0.925 NE Rc urv= 3500 R= 0.99995 OMC DETECTION SYSTEM N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna PR reflection inside IMC: frequency noise Black: PR misaligned Red: PR aligned N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna PR reflection scattering inside IMC MC scattering (10 ppm) IMC Cavity effect (10% fringes) N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation PR Paolo La Penna Plane Power Recycling mirror The old PR was made by two parts: a curved one inside an external cylindrical glass mass (it was a lens, part of the telescope for collimating inside the ITF) The curved part was fixed to the cylindrical by means of a steel ring and pression screws; There was evidence of mechanical resonances (drifting in time): problems in locking acquisition and in the future Frequency Stabilization: need of a monolitic mirror Decided to make it plane (get rid of transverse movements of the beam induced by the suspension displacements) N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna IMC and RFC alignment 10% ABP OFF ABP dismounted The beam was atomatically aligned on the RFC (fixed below the IB) with ABP The IMC was partly automatically aligned (feedback only on the MC mirror The RFC was more stable than IMC: the AA of the RFC worked quite well the poor AA of the IMC couldn’t correct the drifts N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna PR reflection: Input Beam Attenuation (temporary solution) IMC 10% M6 From the laser Reference Cavity N5-WP1 7TH MEETING, Hannover, 12/12/2005 To the interferometer 700 mW (10% of the full power) The New Virgo Injection Bench installation Paolo La Penna The PR feedback inside the IMC (mid of 2004) Simulations on a lock acquisition technique developed following the LIGO experience Locking trials with this baseline technique failed (first half of July) Beam attenuation installed (summer) N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna HR M6 vs attenuated M6 HR M6 Hz N5-WP1 7TH MEETING, Hannover, 12/12/2005 (Same y-scale) R=10% M6 Hz The New Virgo Injection Bench installation Paolo La Penna The PR feedback inside the IMC (end of 2004) Restart of the locking trials with the baseline technique (21st September) Establishement of theVariable Finesse lock acquisition technique (October) PR locking acquisition after one month (end of October 2004) CONFIRMATION OF THE BACKSCATTERING PROBLEM NEED OF AN OPTICAL ISOLATOR (FARADAY) N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Resume: why to build a new Injection Bench The reflection of the light inside the Input Mode Cleaner increased the frequency noise, thus making it impossible to lock the recycling cavity The use of a Faraday isolator was necessary There was no way to accomodate a Faraday isolator and respective telescope on the old IB: the bench had to be replaced In addition to this: It was decided to replace the curved PR with a plane PR (less coupling with the transverse PR displacement): a short (parabolic) telescope on the IB was necessary It was decided to change the automatic alignment system of the injection system: separate alignment of the Input Mode Cleaner and Reference Cavity N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna How did we proceed The aforementioned problems were individuated in many data takings (C1-C4) The mechanical design of the bench was performed using essentially Autocad and Ansys The optical simulation of the bench was performed using mainly Zemax (simulation of optical aberrations and stray light), and by Optocad The Faraday isolator was tested in Nice with the 20W laser The new IB was mounted and prealigned in Class 10 room (after old IB dismounting) Once prealigned it has been inserted into the tower N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Upper part layout Autocad drawing N5-WP1 7TH MEETING, Hannover, 12/12/2005 Optocad drawing The New Virgo Injection Bench installation Paolo La Penna Lower (RFC) part layout Autocad drawing N5-WP1 7TH MEETING, Hannover, 12/12/2005 Optocad drawing The New Virgo Injection Bench installation Paolo La Penna Autocad 3-D simulation (example) N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna ANSYS model Marionette Reference Cavity N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna ANSYS simulation 30 40 Uy Tx Tz 50 60 70 80 90 Probably Susp. Dihedron Wire Legs couplers (measured in september 2005) N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Plane PR: Beam simulation with spherical mirrors Input mirrors N5-WP1 7TH MEETING, Hannover, 12/12/2005 End mirrors The New Virgo Injection Bench installation Paolo La Penna Plane PR: Beam simulation with parabolic mirrors Input mirrors N5-WP1 7TH MEETING, Hannover, 12/12/2005 End mirrors The New Virgo Injection Bench installation Paolo La Penna Plane PR: (new) 6 parabolic off-axis telescope With plane PR: a big magnification is needed, the telescope has to be short (about 700 mm) Parabolic mirrors are needed The computed curvature radius are not off-shelf: custom mirrors (Optical Surface) IMC: waist 4.9 mm Condensing telescope FI: waist 2.65 mm M5: f= 75 mm M6: f = 600 mm Ø 10 cm N5-WP1 7TH MEETING, Hannover, 12/12/2005 to PR: waist=20 mm d= 675 mm The New Virgo Injection Bench installation Paolo La Penna Parabolic telescope mechanical mount Two open loop picomotors Three closed loop picomotors X and Z closed loop translators Mounts designed by ourselves and Machined by a workshop close to Virgo Mounting commercial actuators N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Alignment of the parabolic telescope (Class 10 room) Laser N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Starting conditions: misaligned, mismatched AUTOCOLLIMATOR N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna 1st autocollimator alignment AUTOCOLLIMATOR REF. MIRROR AUTOCOLLIMATOR N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna 2nd autocollimator alignment AUTOCOLLIMATOR AUTOCOLLIMATOR REF. MIRROR N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Autocollimators aligned AUTOCOLLIMATOR AUTOCOLLIMATOR N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna 1st mirror alignment AUTOCOLLIMATOR AUTOCOLLIMATOR N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna 1st mount alignment AUTOCOLLIMATOR AUTOCOLLIMATOR N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna 2nd mount alignment AUTOCOLLIMATOR AUTOCOLLIMATOR N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna 2nd mount alignment AUTOCOLLIMATOR AUTOCOLLIMATOR N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Mirror matching AUTOCOLLIMATOR AUTOCOLLIMATOR N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Focussing AUTOCOLLIMATOR AUTOCOLLIMATOR N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Off-axis AUTOCOLLIMATOR AUTOCOLLIMATOR N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Off-axis AUTOCOLLIMATOR AUTOCOLLIMATOR N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Off-axis AUTOCOLLIMATOR AUTOCOLLIMATOR N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Autocollimation AUTOCOLLIMATOR AUTOCOLLIMATOR N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Beam dumpers: aluminum boxes and buffle-glass plates N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Beam dumpers and PSDs PSD Beam dumpers PSD PSD PSD N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Resume of the main differences between new and old IB Different shape and strutcture of the bench (octagonal, larger, M6 grid of holes, …) Thinner suspension wires Presence of the Faraday isolator Presence of a collimating (reducing) telescope between the Input Mode Cleaner (two lenses) Presence of a short (less than 1-m-long) off-axis parabolic telescope (×8) Separate alignment of the IMC and the RFC RFC placed after the IMC (no beam on the RFC when the IMC is not locked): it should make the beam on the RFC more stable Use of several PSDs to monitor the beam on the bench Use of larger mirrors Use of specifically designed beam dumpers N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Beam diameter 10 mm N5-WP1 7TH MEETING, Hannover, 12/12/2005 Beam diameter 40 mm The New Virgo Injection Bench installation Paolo La Penna Actions after the recycling cavity locking Decision to revise the Injection Bench design (June 2004) Creation of a group charged to redesign the IB (July 2004) Choice of the Faraday isolator and simulation of several possible telescopes (autumn 2004) Decision to install a plane PR (and therefore an off-axis parabolic telescope) (November 2004) Develope of the IB optical and mechanical design and starting of orders and (first months of 2005) Arrival of the main part of the components (May 2005) Arrival of the bench (July 2005) Arrival (delayed) of the last parabolic mirror (M5, 75 mm focal length) (August 2005) N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Actions after the recycling cavity locking Dismounting of the old IB and starting of the mounting of the new IB (September 2005) Insertion of the new IB into the IB tower (begin of november 2005) IB suspension tuning (november 2005) Bench prealignment and local controls tuning (end of November 2005) Mounting of the plane PR (end of November 2005) Local control of the bench (December 2005) Nd:YAG injection and beam alignment (December 2005) Alignment of the IB optics (now) N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Brewster dielectric polarizers Rotator February 2005 N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Mid September 2005 N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Begin of October 2005 N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Main activities before inserting the bench into the tower Change of the input flange (larger windows) Change of the Brewster link (to accomodate a larger beam) Suspension tuning (filter 0 and 7 blades): the bench is heavier (about 150 kg) Adjustment of the wires length Full tuning and balancing of the suspension Mounting of the coils: close to the center of mass plane (more stable equilibrium of the bench) Cabling (of coils and bench actuators) N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Insertion into the tower (begin November) Detector table N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Insertion into the tower (begin November) Detector table N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Begin of November 2005 N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Now N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Now N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Main activities after inserting the bench into the tower Bench heigth setting: some difficulties with the suspension point motors Balancing of the bench Orientation of the bench with an HeNe coming from the MC tower: some difficulties in the filter 7 rotation range Test of the coils: some problems in the frames interrupting RFC beams Orientation of the local controls target: quite difficult because the small angular range in its orientation N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Local controls The IB local controls use a target, placed in front of the dihedron, with reference markers (bench translations) and reflecting HeNe beams (angles) The setup had not been changed with respect to the old IB Beam diameter 10 mm N5-WP1 7TH MEETING, Hannover, 12/12/2005 Beam diameter 40 mm The New Virgo Injection Bench installation Paolo La Penna Old IB local controls N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna New IB local controls After the realignment of the reference beams and target: Closing of the coarse and fine with the old IB correction filters and essentially the same gains (end of November) Adjustment of the reference spots Fine alignment of the bench with the MC HeNe (begin of December) Measurement of the mechanical TF (last week) N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Preliminary: New TF measurement (Henrich Heitmann and Paolo Ruggi) N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Full alignment procedure After this first alignment in Class 100 room, the bench has been mounted inside the tower. The dummy dihedron is mounted, the LC target is realigned, the local controls of the bench are restarted. The IB is aligned using the HeNe coming from the MC tower The Nd:YAG beam is injected into the bench and aligned until a light spot is visible on the MC mirror The optics on the bench is realigned using the IB actuators The true dihedron is reinstalled, the IB is put in vacuum, the IMC aligned and locked After IMC locking, it should be possible to make the final tuning and alignment using the bench actuators. N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna Next steps Present status: The fake dihedron is still mounted on the bench The Nd:YAG beam (up to 5 W for the moment) has been sent into the bench: some light (with even some interference) is visible at the end of the MC The optics on the bench are being realigned using the IB actuators: we perform this operation staying inside the tower, with the laminar flux on and the LC closed (they are very robust) The true dihedron will be reinstalled this week, then the IMC will be aligned (we hope it will be possible in air, in order to be able check the RFC alignment) the IB will put in vacuum, the IMC aligned and locked After IMC locking, it should be possible to make the final tuning and alignment using the bench actuators. The beam will be aligned in the ITF N5-WP1 7TH MEETING, Hannover, 12/12/2005 The New Virgo Injection Bench installation Paolo La Penna